Gas warning unit for chlorine, chlorine dioxide, ozone, ammonia and hydrochloric acid
Optimum safety – minimum labour input: The Conex DIA-G safety system monitors the concentration of the gases chlorine, chlorine dioxide, ozone, ammonia and hydrochloric acid in the air of storage rooms. It is equipped with a range of functions providing the users with optimum safety combined with minimum labour input. One unit can use two interconnected sensors to monitor two different gases, two different gas concentrations or two different gas storage rooms in parallel.


Conex DIA-G with two amperometric sensors

FOR SAFE MONITORING OF
Chlorine
Chlorine Dioxide
Ozone
Ammonia
Hypochloric Acid


Conex DIA-G with one amperometric sensor and one potentiostatic sensor (plugged into a Sensor interface)


Conex DIA-G with two potentiostatic sensors (plugged into two Sensor interfaces)

AMPLIFIER CONEX DIA-G
Electronics
16 bit microprocessor
Relay outputs
5 potential-free relay outputs per software switchable to N.O. or N.C. (fail-safe), max. 250 V/6 A, max. 550 VA
  • 2 relays for the limit values of every of the two sensors
  • 1 alarm relay
Power supply
110 – 240 V (50/60 Hz) or 24 V DC
Separate Sensor Interface
For one potentiostatic sensor, directly plugged into the Interface



GASESMEASURING RANGEACCURACYRESPONSE TIME
(20°C)
MAX. DISTANCE FROM AMPLIFIER
Amperometric sensors
Cl2
0 – 5 ppm
+/-10%
2 s
100 m
ClO2
0 – 5 ppm
+/-10%
2 s
100 m
O3
0 – 5 ppm
+/-10%
2 s
100 m
Potentiostatic sensors
Cl2
0.00 – 20.00 ppm
< 0.05 ppm
< 30 s
500 m
ClO2
0.00 – 1.00 ppm
< 0.03 ppm
< 120 s
500 m
O3
0.00 – 1.00 ppm
< 0.02 ppm
< 60 s
500 m
NH3
0 – 100 ppm
< 1 ppm
< 120 s
500 m
HCl
0.00 – 30.0 ppm
< 0.07 ppm
< 70 s
500 m



FEATURES AND BENEFITS
Optimum safety: Thanks to permanent sensor monitoring, the alarm relay and the possibility of backup operation (uninterrupted current supply) by connection of an external buffer battery
Very short response time in case of a sudden change of the gas concentration
Long service life of the sensor element – without necessity of maintenance
Sensor recognition and auto-calibration, monitoring of the lifetime of the sensor
Internal CAN bus interface for the connection of potentiostatic sensors using bus lines with a total length of max. 500 m
External CAN bus interface for the connection of Conex DIA-G to the visualising software AquaVision 2 or to PLC